R. Dickerhoff et A. Vonruecker, ENUMERATION OF PLATELETS BY MULTIPARAMETER FLOW-CYTOMETRY USING PLATELET-SPECIFIC ANTIBODIES AND FLUORESCENT REFERENCE PARTICLES, Clinical and laboratory haematology, 17(2), 1995, pp. 163-172
The correct enumeration of platelets is still an elusive matter. This
is mainly due to the fact that commercial instruments which are used f
or platelet counting cannot discriminate platelets from other cellular
particles and precipit;ates that cause similar signals. Visual (chamb
er counting) methods are still frequently use:d in routine laboratorie
s to verify low automated platelet counts (< 50 x 10/1) despite obviou
s technical and statistical drawbacks. The following report shows how
platelet counts call be measured by multiparameter now cytometry with
the help of reference particles (fluorescent latex beads) and platelet
-specific antibodies i,e. anti-GPIIb/IIIa(CD41a), anti-GP Ib-alpha (CD
42b) and anti-GP IIIa (CD61), The linearity of this method was highly
satisfactory anti the observed imprecision was within acceptable limit
s. At a platelet concentration of 10 x 10(9)/1 the coefficient of vari
ation (CV, n = 10) ranged from 5.3% (PCV = 0.456) to 5.6% (PCV = 0,148
), Accuracy was evaluated by comparing results to the ICSH-selected me
thod for platelet counting. The correlation of both methods was signif
icant (P < 0.005) and Passing-Bablok's linear regression analysis show
ed no systematic differences between the two methods, Comparisons of t
his new platelet counting technique were also performed with routine v
isual methods, automated blood analysers (Technicon H-l, Sysmex E-5000
) and a different flow cytometric method using only forward and side l
ight scatter properties of platelets for their discrimination, The lin
ear correlation of all methods was significant (P < 0.01) at platelet
concentrations above 50 x 10(9)/1. At lower platelet concentrations, o
ur new platelet counting technique correlated significantly only with
the visual and the forward/side scatter methods. These findings stress
the necessity to confirm low platelet counts by automated blood analy
sers and suggest that using multiparameter now cytometry with platelet
-specific antibodies may be a proficient way to do so, The possibility
of using this technique as a reference method is discussed.