Em. Fernandez et al., THE ACCEPTABILITY OF A MULTILEAF COLLIMATOR AS A REPLACEMENT FOR CONVENTIONAL BLOCKS, Radiotherapy and oncology, 36(1), 1995, pp. 65-74
Citations number
16
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
A multileaf collimator (MLC) can be used as a replacement for conventi
onal blocks as well as for conformal radiotherapy. This study has asse
ssed the possibility of using a Philips MLC for 218 patients treated w
ith conventionally blocked fields. It was found that MLC field shaping
would have been appropriate for over 94% of such patients. The facili
ty to treat large blocked fields has been found to be particularly use
ful. Use of the predefined shapes stored in the Regular Shape Library
provided by Philips was evaluated and it was found that an appropriate
shape was available in 52% of cases. The application of MLC fields to
the treatment of different anatomical sites is discussed.