RADIOTHERAPY TREATMENT PLANNING WITH DYNAMIC WEDGES - AN ALGORITHM FOR GENERATING WEDGE FACTORS AND BEAM DATA

Citation
Sj. Thomas et Kr. Foster, RADIOTHERAPY TREATMENT PLANNING WITH DYNAMIC WEDGES - AN ALGORITHM FOR GENERATING WEDGE FACTORS AND BEAM DATA, Physics in medicine and biology, 40(9), 1995, pp. 1421-1433
Citations number
6
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
40
Issue
9
Year of publication
1995
Pages
1421 - 1433
Database
ISI
SICI code
0031-9155(1995)40:9<1421:RTPWDW>2.0.ZU;2-W
Abstract
If the jaws of a linear accelerator are moved under computer control d uring irradiation, dose distributions similar to those with wedge filt ers can be produced. Varian linear accelerators utilize this effect to give a 'dynamic wedge', using segmented treatment tables (STTs). An a lgorithm is described to generate the dose per monitor unit at any poi nt in a beam, using the STT values. Dynamically wedged beams are model led as the superposition of static asymmetric beams, using an algorith m based on beam data measured for symmetric beams. Predictions of wedg e factors, depth doses and profiles generated using the algorithm are compared with measurements. Good agreement is found between prediction s and measurements. The calculation time is typically 5 ms/dose point on a PC with a 486DX processor.