Mj. Lock et al., DEVELOPMENT OF A QUANTITATIVE COMPETITIVE POLYMERASE CHAIN-REACTION FOR HUMAN HERPESVIRUS-8, Journal of virological methods, 64(1), 1997, pp. 19-26
Citations number
26
Categorie Soggetti
Virology,"Biochemical Research Methods","Biothechnology & Applied Migrobiology
A quantitative competitive polymerase chain reaction (PCR) assay for h
uman herpesvirus 8 (HHV8) was developed. The assay uses a competitive
internal control sequence which differs from the wild type sequence by
the presence of an EcoRI restriction endonuclease site. The quantitat
ive method was validated by co-amplifying known amounts of control seq
uence DNA and wild type DNA, and shown to accurately quantify HHV8 wit
hin the range of 10(1)-10(6) genome copies (R = 0.998). The assay was
used to quantify HHV8 in sequential blood samples of a renal transplan
t patient diagnosed with Kaposi's sarcoma (KS). A peak viral load of 2
8 320 genomes/ml blood was present at diagnosis of KS, which reduced t
o 13 680 genomes after 2 days and was undetectable 4 months later. Con
trol of HHV8 replication in this patient parallelled the cessation of
immunosuppressive therapy. Copyright (C) 1997 Elsevier Science B.V.