We report a high-performance liquid chromatographic method with column
switching for urinary hypoxanthine and xanthine. Analyses were carrie
d out with both a reversed-phase column and an anion-exchange column c
onnected by a column switch and controlled automatically by a computer
ized system controller. The relationships between standard concentrati
ons and peak heights were linear in a concentration range of 1 to 1000
nmol/ml. The recovery of hypoxanthine added to urine was 101.1%, and
that of xanthine was 98.1%. With our method urinary hypoxanthine and x
anthine can be measured accurately without any sample preparation othe
r than filtration.