J. Hu et al., A FAST, RELIABLE, AUTOMATIC SHIMMING PROCEDURE USING H-1 CHEMICAL-SHIFT-IMAGING SPECTROSCOPY, Journal of magnetic resonance. Series B, 108(3), 1995, pp. 213-219
The importance and the difficulty of achieving good Be homogeneity ove
r the volume of interest in in vivo NMR spectroscopy are well known A
fast, reliable, versatile, and fully automatic shimming procedure has
been developed, using 3D chemical-shift imaging to measure the field d
istribution using only the water peak in the sample of interest, The p
rocedure minimizes the mean-square error in the field distribution wit
h respect to a constant held where the signal exists (e,g., the head).
This produces a set of linear equations that can be solved by standar
d matrix methods. The procedure has been applied on a commercial image
r producing water linewidths from the entire head as low as 8 Hz at 1.
5 T. The mean linewidth and standard deviation from 94 head studies we
re 11.7 +/- 1.9 Hz. (C) 1995 Academic Press, Inc.