Mr. Stringer et al., THE CALIBRATION OF FIBER OPTIC PROBES USED FOR LIGHT DOSIMETRY MEASUREMENTS DURING PHOTODYNAMIC THERAPY, Lasers in medical science, 10(1), 1995, pp. 19-24
We describe the calibration of fibre optic probes used to perform in v
ivo light dosimetry studies during the treatment of skin lesions by ph
otodynamic therapy. Results from six individual detectors show that th
e linearity of the calibration and the calculated radiant energy fluen
ce rate within a liquid phantom are independent of the type and sensit
ivity of the probe. The method of calibration is also shown to yield t
he optical interaction coefficients of the phantom.