A method is introduced to calculate monitor units to points off axis.
Extensive data are presented comparing this method with measured value
s of dose per monitor unit on the central ray of asymmetric fields pro
duced by a variety of linear accelerators. The technique demonstrates
improvement over existing methods that use large-field profile data. T
he method is found to be both simple and accurate: Agreement within +/
-2% is obtained using parameters readily available within the clinic.