H. Gulce et al., AMPEROMETRIC ENZYME ELECTRODE FOR AEROBIC GLUCOSE MONITORING PREPAREDBY GLUCOSE-OXIDASE IMMOBILIZED IN POLY(VINYLFERROCENIUM), Journal of electroanalytical chemistry [1992], 394(1-2), 1995, pp. 63-70
A simple, sensitive, stable and low-cost glucose sensor with a fast re
sponse that operates under aerobic solution conditions is described. T
he electrode consists of a polyvinylferrocenium-perchlorate-coated Pt
surface onto which the enzyme is attached. The current response of the
electrode was determined by measuring either cyclic voltammetric peak
current values or steady state current values obtained after a consta
nt-potential step. The effects of glucose concentration, the amount of
enzyme immobilized, the thickness of the polymeric layer and the oper
ating temperature on the response of the enzyme electrode were studied
. The effects of two common physiological sources of interference were
also investigated. A mechanism for the operation of the electrode is
also proposed.