O. Haupt et al., PREPARATION OF QUARTZ FIBER FILTER STANDARDS FOR X-RAY-FLUORESCENCE ANALYSIS OF AEROSOL SAMPLES, X-ray spectrometry, 24(5), 1995, pp. 267-275
Results are presented for an intercomparison of two different techniqu
es to produce standard filter materials for the energy-dispersive x-ra
y analysis of aerosol filter samples, In one, multi-element standard s
olutions were dropped on to Munktell quartz-fibre filters and air drie
d. The other technique utilized an aerosol generator. The dried aeroso
l was deposited on the filter surface, Extensive calibration procedure
s were performed with both types of standards. The data were applied t
o the analysis of urban aerosol filter samples. The results indicated
that the aerosol technique provides good accuracy even for elements of
atomic numbers below 30. The dropping technique is not applicable far
fibre filters and the calibration of light elements (atomic number <3
0) and achieves lower accuracy. Therefore, the aerosol technique shoul
d be applied generally to the preparation of standard filters. This me
thod provides small particles and homogeneous coatings of the multi-el
ement standards.