Wa. Berg et al., SINGLE-LUMEN AND DOUBLE-LUMEN SILICONE BREAST IMPLANT INTEGRITY - PROSPECTIVE EVALUATION OF MR AND US CRITERIA, Radiology, 197(1), 1995, pp. 45-52
PURPOSE: To evaluate the accuracy of magnetic resonance (MR) and ultra
sound (US) criteria for breast implant integrity. MATERIALS AND METHOD
S: One hundred twenty-two single-lumen silicone breast implants and 22
bilumen implants were evaluated with surface coil MR imaging and US a
nd surgically removed. MR criteria for implant failure were a collapse
d implant shell (''linguine sign''), foci of silicone outside the shel
l (''noose sign''), and extracapsular gel. US criteria were a collapse
d shell,;low-level echoes within the gel, and ''snowstorm'' echoes of
extracapsular silicone. RESULTS: Among single-lumen implants, MR imagi
ng depicted 39 of 40 ruptures, 14 of 28 with minimal leakage; 49 of 54
intact implants were correctly interpreted. US depicted 26 of 40 rupt
ured implants, four of 28 with minimal leakage, and 30 of 54 intact im
plants. Among bilumen implants, MR imaging depicted four of five impla
nts with rupture of both lumina and nine of 10 as intact; US depicted
one rupture and helped identify two of 10 as intact Mammography accura
tely depicted the status of 29 of 30 bilumen implants with MR imaging
correlation. CONCLUSION: MR imaging depicts implant integrity more acc
urately than US; neither method reliably depicts minimal leakage with
shell collapse. Mammography is useful in screening bilumen implant int
egrity.