WATER ROLLS FOR ELECTRON-IRRADIATION OF THE EAR CANAL

Citation
Wh. Morrison et al., WATER ROLLS FOR ELECTRON-IRRADIATION OF THE EAR CANAL, International journal of radiation oncology, biology, physics, 33(2), 1995, pp. 479-483
Citations number
11
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
03603016
Volume
33
Issue
2
Year of publication
1995
Pages
479 - 483
Database
ISI
SICI code
0360-3016(1995)33:2<479:WRFEOT>2.0.ZU;2-L
Abstract
Purpose: To demonstrate that water bolus in the external ear can decre ase the dose inhomogeneity caused by auricular surface irregularities when the ear is in an electron-beam field. Methods and Materials: Thre e-dimensional (3D) dose distributions with and without water bolus in the external ear were calculated for a representative patient. The ele ctron dose calculations were made using the Hogstrom pencil beam algor ithm as implemented in 3D by Starkschall. To demonstrate the use of wa ter bolus in the ear clinically, the case of a patient with squamous c arcinoma of the concha who was treated with electrons is presented. Re sults: Water bolus markedly lessens the dose heterogeneity caused by t he surface irregularities of the ear and the air in the external audit ory canal. In the test case, the maximum dose was reduced by 25% using this technique. Conclusion: When the ear is in an electron beam field , warm water should be placed in the external auditory canal and conch a. This maneuver may reduce the incidence of auricular complications t hat occur after electron-beam therapy.