Ew. Hsu et al., A MODIFIED IMAGING SEQUENCE FOR ACCURATE T-2 MEASUREMENTS USING NMR MICROSCOPY, Journal of magnetic resonance. Series B, 109(1), 1995, pp. 66-69
A modified spin-echo pulse sequence is described that enables accurate
T-2 measurements to be made in NMR microimaging experiments. The modi
fied sequence eliminates cumulative diffusion losses that lead to an u
nderestimation of the T-2 relaxation time using conventional spin-echo
pulse sequences. The approach is theoretically justified and confirme
d in comparative experiments on phantoms. (C) 1995 Academic Press, Inc
.