R. Abe et al., DEVELOPMENT OF AN APPARATUS TO MEASURE PLASMA COLLOID OSMOTIC-PRESSURE MOUNTED ON THE EXTRACORPOREAL HEMODIALYSIS CIRCUIT, Tohoku Journal of Experimental Medicine, 176(3), 1995, pp. 149-154
We newly developed an apparatus to measure colloid osmotic pressure (C
OP) of blood plasma continuously by hemodialysis (HD) or extracorporea
l ultrafiltration method (ECUM) without any invasive procedure or bloo
d sampling. The COP value measured by this apparatus highly correlated
with the COP value calculated from the concentration of total protein
of blood plasma (TP) sampled simultaneously (y=21.182+0.96256x, r(2)=
0.911). Measurement of the COP with this apparatus seems to be reliabl
e.