T. Ilchmann et al., MEASUREMENT ACCURACY IN ACETABULAR CUP WEAR - 3 RETROSPECTIVE METHODSCOMPARED WITH ROENTGEN STEREOPHOTOGRAMMETRY, The Journal of arthroplasty, 10(5), 1995, pp. 636-642
The accuracy of three methods (the simple and noncomputerized Scheier-
Sandel and Charnley-Duo methods and the computerized fin Bild Roentgen
Analyse [EBRA] method) for retrospective wear measurements of the ace
tabular cup from standard pelvis radiographs was studied. Measurements
on 13 hip prostheses were compared with those obtained by roentgen st
ereophotogrammetry analysis. The Scheier-Sandel method had the lowest
accuracy and the EBRA method had the best accuracy. The Charnley-Duo m
ethod was almost as good when starting analysis 3 months after surgery
and is easier to use. The EBRA method is useful for accurate measurem
ents on a small number of patients; the Charnley-Duo method is recomme
nded for clinical wear studies on a larger number of patients.