Sn. Muddukrishna et al., QUANTITATION OF REDUCED DISULFIDE GROUPS IN MONOCLONAL-ANTIBODIES USING 5-IODOACETAMIDOFLUORESCEIN - A NOVEL SIZE EXCLUSION-HPLC TECHNIQUE, Applied radiation and isotopes, 46(10), 1995, pp. 1015-1026
Citations number
22
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging
The interference of stannous in the quantitation of sulfhydryl in pret
reated MAb using the Ellman's method is described. A HPLC method which
effectively overcomes the interference by stannous ion in measuring s
ulfhydryls by the commonly used Ellman's technique has been developed.
The method involves reacting the reduced monoclonal antibody with 0.1
M EDTA followed by the sulfhydryl specific reagent 5-iodoacetamidoflu
orescein and incubating the mixture at 37 degrees C for 1.0-1.5 h. The
5-IAF labeled protein is separated from the unbound free 5-IAF and it
s secondary degraded products by size exclusion HPLC using a TSK G3000
SWXL column. The method allows quantitation of MAb sulfhydryl present
as monomeric MAb, aggregates and fragments and is a major improvement
over the spectroscopic techniques.