A flow-injection method was applied to on-line analysis for ammonia in
industrial liquid effluents. The method utilised the diffusion of gas
eous ammonia from an alkaline sample stream, through a microporous mem
brane, into a bromothymol blue indicator stream, with the resulting co
lour change measured photometrically using a red LED. Following initia
l development work, the method was adapted to an automated, portable s
ystem, which was capable of linear response in the range 1-100 mg l(-1
) NH3-N. The portable flow injection monitor was powered by an externa
l 12 V battery, and performed two-point self-calibration during each a
nalytical cycle. The monitor was capable of unattended, continuous on-
line monitoring of effluent streams at two industrial sites over sever
al days. The results obtained were in good agreement with those produc
ed by an indophenol blue laboratory reference method.