Oesophageal malposition of an endotracheal tube is among the leading c
auses of anaesthesia incidents. While clinical manoeuvres for detectio
n of tube malposition are unreliable, monitoring (i.e. capnography) ca
n prevent such incidents. The problem is particularly important in pre
hospital care, where capnography is not (yet) widely available. We tes
ted three devices used for differentiating oesophageal from endotrache
al intubation: 1. Non-CO2-dependent Oesophageal Detector Device (ODD)
as described by Pollard and Wee, 2. Semi-quantitative chemical disposa
ble capnometer EasyCAP (Nellcor), 3. Non-quantitative infrared miniatu
rised capnometer MiniCAP (MSA). Methods. 50 anaesthetised minipigs wer
e intubated with a Magill tube. An identical additional tube was place
d in the oesophagus. The cuffs of both tubes were inflated. Unexperien
ced personel (students, laborary technicians) were asked to determine
the position of one of the tubes by using one of the devices according
to the randomisation plan. The decision had to be taken within 30 s.
Using the ODD, the proband first injected 100 ml air into the lung (or
stomach) and then tried to aspirate the same volume. EasyCAP and Mini
CAP were used according to manuals. Results. Each device was used 25 t
imes with a tracheal tube and 25 times with an oesophageal tube. All t
ube position identifications were correct. When ventilating the oesoph
agus/stomach for capnometric control, regurgitation into the tube occu
rred six times (five times with the EasyCAP and once with the MiniCAP)
. In these cases, the decision was based on this occurrence and not on
the display of the device. While using the ODD no regurgitation occur
red. Conclusion. These devices are useful for preclinical practice. Ac
cording to the literature and our experience, the ODD is superior for
the initial control of tube position, especially in cardiac arrest. Ca
pnometry is needed, however, for continuous control of ventilation.