H. Schuhmacher, ON THE ACHIEVEMENT OF OPERATIONAL RADIATION PROTECTION QUANTITIES FORMIXED PHOTON-NEUTRON RADIATION-FIELDS, Radiation protection dosimetry, 61(1-3), 1995, pp. 63-66
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
In calibration fields for penetrating radiation, operational radiation
protection quantitites such as H(10) and H-p(10) are commonly derive
d from measurable quantities (spectral fluence for neutrons,, air kerm
a for photons) and calculated fluence to dose equivalent coefficients
for neutrons and kerma to dose equivalent conversion coefficients for
photons. The paper presents a theoretical study on the suitability of
a small tissue-equivalent proportional counter (TEPC), inserted at a d
epth of 10 mm in a phantom, for measuring H(10) and H-p(10). The diff
erences between the definition of dose equivalent and the measurement
principle of the TEPC are discussed. If the pressure of the TEPC gas f
illing, the phantom material and size are all correctly chosen, a read
ing which deviates by less than +/-25% from H-p(10) for photons and ne
utrons impinging on the phantom from the entire front hemisphere can b
e obtained.