TEPC MEASUREMENTS WITH THE VARIANCE COVARIANCE METHOD ON BOARD AIRCRAFT

Citation
L. Lindborg et al., TEPC MEASUREMENTS WITH THE VARIANCE COVARIANCE METHOD ON BOARD AIRCRAFT, Radiation protection dosimetry, 61(1-3), 1995, pp. 119-124
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
61
Issue
1-3
Year of publication
1995
Pages
119 - 124
Database
ISI
SICI code
0144-8420(1995)61:1-3<119:TMWTVC>2.0.ZU;2-G
Abstract
A portable instrument based on the variance-covariance method is descr ibed. It consists of two TEPC detectors, which are connected to electr ometers. The electric current and its variance as well as the covarian ce between the two currents are used to determine the dose and the qua lity factor. The paper gives the essential relations from which the qu antities are calculated and presents the main sources of uncertainties . The instrument has been used on board an aircraft at 11 km height fo r measurement of the ambient dose equivalent rate and the average qual ity factor. After normalisation to a reference neutron field at CERN t he ambient dose equivalent rate became (4.5 +/- 1.0) mu Sv.h(-1). For the normalisation it wits important to have an accurate estimate of th e low LET dose fraction.