W. Rosenstock et al., ESTIMATION OF NEUTRON ENERGY-SPECTRA WITH BUBBLE DETECTORS - POTENTIAL AND LIMITATIONS, Radiation protection dosimetry, 61(1-3), 1995, pp. 133-136
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
The capabilities of neutron bubble detectors to give limited energy in
formation have been investigated with neutrons from Cf-252 and D-T sou
rces. The Bubble Detector Spectrometer (BDS) with six different energy
thresholds ranging from 10 keV to 10 MeV has been used. The data obta
ined with the Cf-252 source have been compared with data from measurem
ents with the neutron spectrometer ROSPEC. In addition the data have b
een compared with theoretical calculations. The agreement between the
measured BDS data with the californium source and theory was acceptabl
e. For the measurements at the 14 MeV D-T neutron generator a deconvol
ution of the bubble data has not been possible, therefore the doses ca
lculated from the number of bubbles produced in the detectors are give
n. There is a clear indication of neutrons with energies higher than 1
0 MeV. For these neutron energies the fluence is measured correct with
in an error of 15%.