ESTIMATION OF NEUTRON ENERGY-SPECTRA WITH BUBBLE DETECTORS - POTENTIAL AND LIMITATIONS

Citation
W. Rosenstock et al., ESTIMATION OF NEUTRON ENERGY-SPECTRA WITH BUBBLE DETECTORS - POTENTIAL AND LIMITATIONS, Radiation protection dosimetry, 61(1-3), 1995, pp. 133-136
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
61
Issue
1-3
Year of publication
1995
Pages
133 - 136
Database
ISI
SICI code
0144-8420(1995)61:1-3<133:EONEWB>2.0.ZU;2-7
Abstract
The capabilities of neutron bubble detectors to give limited energy in formation have been investigated with neutrons from Cf-252 and D-T sou rces. The Bubble Detector Spectrometer (BDS) with six different energy thresholds ranging from 10 keV to 10 MeV has been used. The data obta ined with the Cf-252 source have been compared with data from measurem ents with the neutron spectrometer ROSPEC. In addition the data have b een compared with theoretical calculations. The agreement between the measured BDS data with the californium source and theory was acceptabl e. For the measurements at the 14 MeV D-T neutron generator a deconvol ution of the bubble data has not been possible, therefore the doses ca lculated from the number of bubbles produced in the detectors are give n. There is a clear indication of neutrons with energies higher than 1 0 MeV. For these neutron energies the fluence is measured correct with in an error of 15%.