The Axial Separation Module (ASM(TM)), which separates whole-blood spe
cimens serially in Axial Process Containers (APC(TM)), was evaluated f
or clinical performance at the University of Virginia Health Sciences
Center (UVA HSC) in a community-based outpatient laboratory (North Rid
ge Clinic). We hypothesized that moving the task of blood separation t
o point of care would reduce specimen turnaround time within the main
laboratory. Blood drawn into an APC was separated in the ASM at point
of care at the North Ridge Clinic. Blood drawn into a Vacutainer Tube(
TM) was separated in a conventional centrifuge at the main laboratory.
Turnaround time was calculated for the ''chem 17'' test from files st
ored in our laboratory information system. Blood serially separated at
point of care yielded turnaround time savings for specimens originati
ng from the North Ridge Clinic. Average turnaround time decreased by 2
4%. Phlebotomists found no appreciable workload increase from incorpor
ating the ASM as a point-of-care blood separation device. Phlebotomist
s also found that they could immediately detect hemolysis. We conclude
d that serial separation at point of care reduces specimen turnaround
time at the main laboratory. The ASM/APC was found to be better suited
for point-of-care blood separation than a conventional centrifuge. We
speculate that immediate blood separation has the potential to improv
e the quality of analytical results.