TRANSCATHETER CLOSURE OF OVERSIZED PERSISTENT DUCTUS-ARTERIOSUS BY SIMULTANEOUS DELIVERY OF 2 RASHKIND UMBRELLA DEVICES

Citation
H. Sievert et al., TRANSCATHETER CLOSURE OF OVERSIZED PERSISTENT DUCTUS-ARTERIOSUS BY SIMULTANEOUS DELIVERY OF 2 RASHKIND UMBRELLA DEVICES, Catheterization and cardiovascular diagnosis, 36(3), 1995, pp. 251-254
Citations number
33
Categorie Soggetti
Cardiac & Cardiovascular System
ISSN journal
00986569
Volume
36
Issue
3
Year of publication
1995
Pages
251 - 254
Database
ISI
SICI code
0098-6569(1995)36:3<251:TCOOPD>2.0.ZU;2-2
Abstract
Transcatheter closure of persistent ductus arteriosus (PDA) with a dia meter of >9 mm is considered to be impossible or at least difficult wi th the occlusion systems that are currently available. We report a sim ple technique for occluding oversized PDAs with two Rashkind umbrellas introduced simultaneously, Using this technique, a large PDA with a d iameter of 13 mm was successfully occluded in a 40-year-old man. Compl ete closure without residual shunt was documented by echocardiogram an d angiogram. (C) 1995 Wiley-Liss, Inc.