Ds. Srivastava et al., CALIBRATION FACTOR FOR LR-115 TYPE-II TRACK DETECTORS FOR ENVIRONMENTAL RADON MEASUREMENTS, Nuclear geophysics, 9(5), 1995, pp. 487-495
Citations number
NO
Categorie Soggetti
Geosciences, Interdisciplinary","Mining & Mineral Processing","Nuclear Sciences & Tecnology
This paper presents the experimental calibration of LR-115 type II pla
stic track detectors for monitoring environmental radon. The detectors
were calibrated using a standard radon chamber and simulating the env
ironmental conditions as normally found in dwellings. The value of the
calibration factor was found to be 625 tracks cm(-2) d(-1) per WL in
the BARE mode and 0.116 tracks cm(-2) d(-1) per Bq m(-3) for cup with
membrane mode. Results of measurements of radon and its daughters in B
ijnor (India) are also presented.