An immunoassay technique based on a high surface area carbon immunoele
ctrode is described. Dispersed ULTI carbon serves as a carrier for imm
obilized antibodies and also as an electrode material. 'Sandwich' sche
me of immunoassay of rabbit IgG (as a model analyte) has been used. Io
dine formed as a result of the enzymatic oxidation of iodide by peroxi
dase-label has been detected amperometrically. Using dispersed carbon
material as a solid support for immobilization of immunoagents and as
an electrode material improves the efficiency of immune-interaction an
d the sensitivity of electrochemical assay due to the high area-to-vol
ume ratio of solid to liquid phases. The method based on 'sandwich' im
munoassay is conducted at 21 degrees C. The time of each stage of incu
bation (with analyte and with peroxidase-antibody conjugate) is 10 min
. Electrochemical detection of peroxidase-labeled immuno-complex does
not exceed several minutes. The technique allows fast determination of
rabbit IgG with a low detection limit, in the nanomolar range. By com
parison, the conventional technique based on ELISA, being performed un
der the same conditions (temperature and incubation time) does not all
ow any quantitative determination of IgG for concentration lower then
100 nM.