Pm. Sapin et al., THE ASSESSMENT OF LEFT-VENTRICULAR FILLING DYNAMICS USING AN ONLINE AUTOMATIC BORDER DETECTION ALGORITHM - COMPARISON WITH CINEVENTRICULOGRAPHY, Echocardiography, 12(6), 1995, pp. 559-569
An echocardiographic system has been developed that performs automatic
endocardial border detection. and instantaneously calculates and disp
lays a waveform of left ventricular cavity area versus time. The purpo
se of this study was to compare measurements of left ventricular filli
ng dynamics from automatic border detection. echocardiography with sim
ilar measurements from cineventriculography. Thirty-three patients und
ergoing cardiac catheterization. had automatic border detection echoca
rdiography performed within 45 minutes of cineventriculography. Ten pa
tients had normal catheterization findings and 23 had cardiac disease.
The automatic border detection waveforms generated from two echocardi
ographic views were measured to determine the fraction of filling occu
rring during the early diastolic rapid filling phase and during the fi
lling phase resulting from atrial contraction. Similar fractions were
derived from curves generated from frame-by-frame measurements of cine
angiographic volumes. Results were analyzed by correlating echocardiog
raphic and cineventriculographic results, and by a (limits of agreemen
t analysis were +/- 2 standard deviations of the mean difference betwe
en echocardiography and cineventriculography). There were significant
correlations between echocardiography and cineventriculography for eac
h of the parameters studied. The best results were obtained for the ap
ical four-chamber view (rapid filling fraction r = 0.72, P < 0.0001, a
trial filling fraction r = 0.56, P < 0.001). Differences in. filling p
atterns between normal and abnormal patient groups detected by cineven
triculography were also detected by automatic border detection. echoca
rdiography. However, broad limits of agreement were observed, that may
limit the ability of the automatic border detection system to reliabl
y predict cineventriculographic results in an individual patient. Auto
matic border detection echocardiography can provide information about
left ventricular filling dynamics that is similar to that obtained fro
m frame-by-frame analysis of cineventriculograms. However the variabil
ity in the results may limit the application of the technique in indiv
idual patients.