SULFATE PRECONCENTRATION BY ANION-EXCHANGE RESIN IN FLOW-INJECTION AND ITS TURBIDIMETRIC DETERMINATION IN WATER

Citation
Mms. Filha et al., SULFATE PRECONCENTRATION BY ANION-EXCHANGE RESIN IN FLOW-INJECTION AND ITS TURBIDIMETRIC DETERMINATION IN WATER, Talanta, 40(10), 1993, pp. 1529-1534
Citations number
8
Journal title
Talanta
ISSN journal
00399140 → ACNP
Volume
40
Issue
10
Year of publication
1993
Pages
1529 - 1534
Database
ISI
SICI code
0039-9140(1993)40:10<1529:SPBARI>2.0.ZU;2-J
Abstract
A preconcentration procedure was established for sulphate determinatio n in rain waters at the mg/l level, employing a small column packed wi th the AG1-X8 (200-400 mesh) anionic resin inserted into a flow inject ion system. Sulphate determination was performed by using the turbidim etric method based on reaction with barium. For concentrations within 0.10 and 2.0 mgSO42-/l, a throughput of 50 determinations/hr was achie ved, and the relative standard deviation of results was better than 2% .