A fault in the assembly of a Matrix Large Animal Circle anaesthetic ma
chine resulted in reversal of fresh gas flow through the vaporizer. Th
e fault was discovered only after the sudden development of excessive
depth of anaesthesia in two equine patients. Laboratory investigations
were conducted to determine the effect of flow reversal on vaporizer
output. Results indicated that output concentration was approximately
doubled under these conditions.