FLOW REVERSAL THROUGH A MARK-III HALOTHANE VAPORIZER

Citation
As. Gregg et al., FLOW REVERSAL THROUGH A MARK-III HALOTHANE VAPORIZER, British Journal of Anaesthesia, 71(2), 1993, pp. 303-304
Citations number
8
Categorie Soggetti
Anesthesiology
ISSN journal
00070912
Volume
71
Issue
2
Year of publication
1993
Pages
303 - 304
Database
ISI
SICI code
0007-0912(1993)71:2<303:FRTAMH>2.0.ZU;2-M
Abstract
A fault in the assembly of a Matrix Large Animal Circle anaesthetic ma chine resulted in reversal of fresh gas flow through the vaporizer. Th e fault was discovered only after the sudden development of excessive depth of anaesthesia in two equine patients. Laboratory investigations were conducted to determine the effect of flow reversal on vaporizer output. Results indicated that output concentration was approximately doubled under these conditions.