A 3-D RADIOSURGICAL METHODOLOGY FOR COMPLEX ARTERIOVENOUS-MALFORMATIONS

Citation
D. Lefkopoulos et al., A 3-D RADIOSURGICAL METHODOLOGY FOR COMPLEX ARTERIOVENOUS-MALFORMATIONS, Radiotherapy and oncology, 28(3), 1993, pp. 233-240
Citations number
18
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
01678140
Volume
28
Issue
3
Year of publication
1993
Pages
233 - 240
Database
ISI
SICI code
0167-8140(1993)28:3<233:A3RMFC>2.0.ZU;2-H
Abstract
A 3-D methodology, the associated targets methodology, for planning ra diosurgical irradiations of complex arteriovenous malformations (AVMs) is presented. It uses the ARTEMIS-3D treatment planning system and ha s been devised and adopted by our group since January 1990. Its main f eatures are: (a) prescription and delivery of a minimal target dose on the surface of the lesion, corresponding to a 60-70% isodose range. T he dose to adjacent functional neurological structures is taken into a ccount as well as the maximum dose to the lesion; (b) An optimisation approach consisting of obtaining the optimal superimposition of the is odose surface and the 3-D contour of the lesion and sharp fall-offs by interactive manipulation of the treatment parameters. The clinical ch oice of the treatment plan is based on a compromise between the optima l reference isodose surface encompassing the lesion and the minimisati on of the volumetric dose fall-off. In complex AVMs the angiographic r esults have been significantly improved in comparison with our previou s experience because of the better achieved lesion encompassing.