QUALITY ASSURANCE IN BRACHYTHERAPY - THE DISPLACEMENT EFFECT IN THE VICINITY OF CO-60 AND IR-192 BRACHYTHERAPY SOURCES

Citation
H. Tolli et Ka. Johansson, QUALITY ASSURANCE IN BRACHYTHERAPY - THE DISPLACEMENT EFFECT IN THE VICINITY OF CO-60 AND IR-192 BRACHYTHERAPY SOURCES, Physics in medicine and biology, 38(10), 1993, pp. 1485-1492
Citations number
7
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
38
Issue
10
Year of publication
1993
Pages
1485 - 1492
Database
ISI
SICI code
0031-9155(1993)38:10<1485:QAIB-T>2.0.ZU;2-C
Abstract
Cylindrical ionization chambers with varying radii have been used to d etermine the displacement effect at distances between 10 mm and 40 mm from Co-60 and Ir-192 brachytherapy sources. Agreement with published values of the displacement factor for external Co-60 beams is obtained . The shift between the centre of the ionization chamber and the effec tive point of measurement is found to be only a few per cent of the in ternal radius of the cylindrical ionization chamber. The shift is foun d to be dependent on the source to ionization chamber centre distance.