An in vitro study was conducted, using a stenotic model to demonstrate
that different pressure gradients can be produced by the same degree
of valvular stenosis. This model is comprised of two cylindric chamber
s with a diaphragm in the center which had a small central hole. An in
jector was connected to one end of the prestenotic chamber to produce
a steady pulsed flow. A rubber tube was connected to the other end of
the poststenotic chamber and led upward to a large reservoir which pro
vided a constant afterload pressure. Two pressure transducers were att
ached to the two connecting tubes, both linked with two pigtail cathet
ers which were accommodated in the chambers just before and after the
stenotic diaphragm. Two sets of injection volumes (20 and 30 ml) and m
ultiple injections with different flow rates (5, 10, 15,...49 ml/s) we
re administered and resulting pressures measured by the two transducer
s were recorded. Results showed that different pressure gradients coul
d be produced using the same injection volume, the same afterload, and
the same degree of stenosis. The greater the flow rate, the higher th
e pressure gradient. Good correlation existed between the pressure gra
dient and the injection flow rate (r = 0.95 and 0.97 for the study gro
ups receiving 20 and 30 ml injection volumes, respectively; p< 0.001 i
n all comparisons). Thus, a higher pressure gradient may not necessari
ly indicate a severe degree of valvular stenosis. Evaluation of a sten
otic lesion should not be made from the degree of pressure gradient al
one-other hemodynamic conditions should also be taken into account.