Motion analysis of the upper extremity during functional activities ha
s only recently become more accessible. An optoelectric system (WATSMA
RT) was subjected to testing using first a calibration dummy and then
human subjects. The mean differences between the system calculated ang
les and those measured with potentiometers on a calibration dummy were
less than 1.6-degrees in all three planes. The test-retest reliabilit
y of the system when measuring elbow motion of human subjects yielded
acceptable repeatability for measurement of functional activities. Cal
culation of the least significant difference found that minimum differ
ences from 4 to 14-degrees in flexion and from 10 to 19-degrees in rot
ation can be detected using the reported set-up and protocol.