ANALYSIS OF IMAGE QUALITY PARAMETERS USING A COMBINATION OF AN ANSI TYPE PHANTOM AND THE LEEDS TOR(CDR) TEST OBJECT IN SIMULATIONS OF SIMPLE EXAMINATIONS

Citation
Al. Llorca et al., ANALYSIS OF IMAGE QUALITY PARAMETERS USING A COMBINATION OF AN ANSI TYPE PHANTOM AND THE LEEDS TOR(CDR) TEST OBJECT IN SIMULATIONS OF SIMPLE EXAMINATIONS, Radiation protection dosimetry, 49(1-3), 1993, pp. 47-49
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
49
Issue
1-3
Year of publication
1993
Pages
47 - 49
Database
ISI
SICI code
0144-8420(1993)49:1-3<47:AOIQPU>2.0.ZU;2-Z
Abstract
Following the criteria of the British Institute of Radiology a global image quality test, employing the Leeds TOR(CDR) test object has been systematically used as a starting point of a QC programme in the Madri d area. In order to guarantee the correlation of the image quality of the phantom with the CEC image quality criteria and other criteria fro m the radiological point of view, some patient equivalent structures ( similar to those described in the ANSI type phantom) have been joined to the Leeds TOR(CDR) test object. A novel use of TOR(CDR) is presente d that attempts to assess image quality under conditions more relevant to some clinical applications. From the results presented, some concl usions are drawn relating to the suitability and accuracy of using mix ed phantoms such as the one described.