MRI MEASUREMENTS OF THE DEPENDENCE ON T-1 OF THE ECHO AMPLITUDES USING A MULTIPLE SPIN-ECHO SCHEME

Citation
A. Fransson et al., MRI MEASUREMENTS OF THE DEPENDENCE ON T-1 OF THE ECHO AMPLITUDES USING A MULTIPLE SPIN-ECHO SCHEME, Magnetic resonance imaging, 11(8), 1993, pp. 1185-1192
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
0730725X
Volume
11
Issue
8
Year of publication
1993
Pages
1185 - 1192
Database
ISI
SICI code
0730-725X(1993)11:8<1185:MMOTDO>2.0.ZU;2-O
Abstract
Analytical calculations using the Bloch formalism were performed to as sess the dependence on T-1 of the echo amplitudes for the Phase-Altern ating Phase-Shift (PHAPS) multiple spin-echo protocol. Measurements in a 0.5 T MR imaging unit were performed to ratify the analytical resul ts. Especially for low T-2 values, the echo amplitudes were erroneous, with an increasing contribution from stimulated echo components with increasing T-1. Apart from affecting T-2 estimates, stimulated echoes generated a non-monoexponential signal decay of the echo trains. The r esults confirmed previous simulation studies as regards the dependence on T-1 of T-2 estimates from PHAPS.