Z. He et al., A RATIO PULSE-SHAPE DISCRIMINATOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 336(1-2), 1993, pp. 236-245
A simple pulse-shape discriminator has been designed to separate signa
ls from a CsI(Na)/GSO phoswich detector. It Can provide information on
both the total energy deposited and the contribution of each crystal
to that total, allowing a software veto to be used according to the di
stribution of the energy deposited. Test results show that even for ph
otons having an energy of 17 keV, the location of most of the interact
ions can be found unambiguously. The system has also been tested on th
e most commonly used NaI(Tl)/CsI(Tl) phoswich detector without particu
lar optimisation of the shaping time-constants. These results show tha
t it can work well for gamma-rays below 17 keV. This technique could b
e very useful in a system which produces more than two decay component
s, such as a triple phoswich detector, or in a system where pulse-shap
e discrimination is used for particle identification.