A STUDY OF INCOMPLETE CHARGE COLLECTION IN CRYOGENIC DETECTORS USING A SEGMENTED 60 GRAM GERMANIUM PHONON AND IONIZATION DETECTOR

Citation
Ba. Young et al., A STUDY OF INCOMPLETE CHARGE COLLECTION IN CRYOGENIC DETECTORS USING A SEGMENTED 60 GRAM GERMANIUM PHONON AND IONIZATION DETECTOR, Journal of low temperature physics, 93(3-4), 1993, pp. 393-398
Citations number
8
Categorie Soggetti
Physics, Applied
ISSN journal
00222291
Volume
93
Issue
3-4
Year of publication
1993
Pages
393 - 398
Database
ISI
SICI code
0022-2291(1993)93:3-4<393:ASOICC>2.0.ZU;2-L
Abstract
We have tested a new 60 gram germanium phonon and ionization detector to study incomplete charge collection effects in cryogenic detectors. An understanding of such effects is necessary to further improve the n uclear vs. electronic recoil discrimination capability of cryogenic de tectors. The detector is operated at almost-equal-to 20 mK. If has thr ee concentric ionization channels, and four electric field shaping str uctures on its sides. Phonons are sensed using NTD Ge thermistors. Fir st results obtained using this device and moveable radioactive sources of Am-241 and Co-57 will be presented