D. Yvon et al., BOLOMETERS DEVELOPMENT, WITH SIMULTANEOUS MEASUREMENT OF HEAT AND IONIZATION SIGNALS, AT SACLAY, Journal of low temperature physics, 93(3-4), 1993, pp. 405-410
We report the latest developments of our work on bolometers. Low noise
readout simultaneously in charge and heat channel was achieved. Tests
were conducted using a 7 g bolometer build in Berkeley. Baseline widt
hs 350 eV (1 sigma, phonon) and 1 keV (1 sigma, ionisation) were measu
red. But the 60 keV phonon width was found to be significantly larger
than the baseline width. Also the amplitude of ionisation signal was f
ound to drift with time. This effect motivated a detailed study of ion
isation detectors at very low temperatures. Behavior of PIN silicon di
odes (Hamamatsu) was investigated from room temperature to 3 K. First
results are shown.