A. Schuchert et Kh. Kuck, EFFECTS OF LOCAL STEROID ELUTION ON CHRONIC PACING THRESHOLD AND IMPEDANCE OF ENDOCARDIAL PACEMAKER LEADS, Cardiology, 83(4), 1993, pp. 240-243
To study the effects of steroid elution on chronic pacing threshold an
d impedance, a 4-mm(2) surface area lead with steroid (030-368, Telect
ronics; n = 9) was compared to the same lead without steroid (030-359,
Telectronics; n = 9) 3 months after implantation. The pulse duration
thresholds of the steroid lead were significantly lower at 0.8, 1.6, 2
.5, and 5.0 V. Impedance measured by telemetry at 0.8, 1.6, 2.5, 5.0 a
nd 8.0 V pulse amplitude, and 0.05, 0.25, 0.5 and 1.0 ms pulse duratio
n was imilar for the two leads. Mean impedance was about 650 Omega and
did not change from 0.8 to 8.0 V. The local steroid elution reduced t
he pacing threshold, but had no influence on the impedance. The stable
impedance indicates that the impedance of this lead was not modified
by pulse-dependent variables and, therefore, can also be correctly det
ermined at pulse amplitudes below 5.0 V.