RETRIEVAL OF A RUPTURED SHAFT OF A PTCA-B ALLOON CATHETER

Citation
A. Utech et al., RETRIEVAL OF A RUPTURED SHAFT OF A PTCA-B ALLOON CATHETER, Zeitschrift fur Kardiologie, 82(11), 1993, pp. 727-728
Citations number
17
Categorie Soggetti
Cardiac & Cardiovascular System
Journal title
ISSN journal
03005860
Volume
82
Issue
11
Year of publication
1993
Pages
727 - 728
Database
ISI
SICI code
0300-5860(1993)82:11<727:ROARSO>2.0.ZU;2-R
Abstract
While attempting angioplasty of a very tight and long circumflex lesio n with a metal-shafted monorail catheter, the flexible part of the bal loon catheter shaft broke off within the guiding catheter. The lost pi ece of catheter was then retrieved from the coronary artery by using a n on-the-wire balloon catheter that was inflated inside the distal end of the guiding catheter.