G. Mazur et A. Pethran, DETECTION OF SPECIFIC IGE IN ISOCYANATE AND PHTHALIC-ANHYDRIDE EXPOSED WORKERS - COMPARISON OF RAST RIA, IMMUNO CAP SYSTEM FEIA, AND MAGIC LITE SQ, Allergy, 48(8), 1993, pp. 627-630
Sera from 94 workers occupationally exposed to isocyanate were tested
by RAST RIA, Immuno CAP FEIA, and Magic Lite SQ (ML) against the aller
gens HDI, MDI, TDI, and phthalic anhydride. Twenty sera showed increas
ed levels of diisocyanate-specific IgE antibodies, and five sera were
phthalic anhydride positive. High total IgE titer was not correlated w
ith positive specific IgE titer (r = 0.47), showing that nonspecific I
gE binding was low. The results of ML and CAP correlated well (r = 0.9
1), but tended to be slightly higher than the results obtained with th
e RAST isotope test. CAP and RAST data also correlated significantly (
r = 0.90), but the correlation between ML and RAST data was lower (r =
0.82). The results of the new in vitro tests, CAP and ML, showed good
reproducibility (5 % CV for CAP and 8% CV for ML). In summary, the CA
P and ML methods were found to be appropriate for routine diagnosis of
specific IgE antibodies