Pulse-radiolysis equipment for studies of radical reaction kinetics in
the gas phase is described. The equipment is built around an accelera
tor of the Febetron type for 800 keV electrons. The analysing light is
pulsed with optical feedback regulation of the intensity to increase
the luminosity ca. 50 times during the measurement. It is equipped wit
h photoelectric detection of transient optical absorptions in the time
interval 1 x 10(-8)-5 x 10(-3) s of radicals in the wavelength range
200-900 mm. The stray-light level is low (<1%) even at 200 nm. The dat
a collection system is computerised and controlled by a PC with hard-
and software for recording transients and for setting the wavelength.
Experimental measurements of the kinetics of decay of the CH3 radical
in SF6 bath gas are in the interval +/- 10 % of the literature value.
The equipment is intended for measurements of reaction kinetics relate
d to combustion and atmospheric chemistry.