An Ivalon-plug (PDA-plug) has been developed for transvenous closure o
f patent ductus arteriosus. It can be introduced into the ductus throu
gh standard venous sheaths with a diameter of 8-16 French. The plug is
fixed in its target position by means of four distal (aortal) and two
proximal (pulmonary) nitinol stnrts. In case of malpositioning the sy
stem can be pulled back into and retrieved through the sheath. In anim
al experiments endothelialization of the implanted material was demons
trated. PDA-closure was attempted in 58 patient; the procedure was suc
cessful in 57 cases; a residual ductal flow was present in eight patie
nts. During a follow-up of up to 24 months there were no complications
.