Jm. Houston et al., COMPARISON OF THE NIST HIGH-ACCURACY CRYOGENIC RADIOMETER AND THE NIST SCALE OF DETECTOR SPECTRAL RESPONSE, Metrologia, 30(4), 1993, pp. 285-290
Two independent methods of measurement were used to determine the abso
lute spectral responsivity and external quantum efficiency of light-tr
apping silicon photodiode packages. These trap packages were calibrate
d first by the NIST High Accuracy Cryogenic Radiometer at laser wavele
ngths of 633 nm and 442 nm. They were also measured in the NIST Spectr
al Comparator Facility with working standards traceable to a 100 % qua
ntum efficient radiometer (QED-200). The two sets of measurements agre
e to better than 0,1 % at 633 nm and 0.25 % at 442 nm.