CHARACTERIZATION OF A HIGH-SENSITIVITY COMPOSITE SILICON BOLOMETER

Citation
G. Eppeldauer et al., CHARACTERIZATION OF A HIGH-SENSITIVITY COMPOSITE SILICON BOLOMETER, Metrologia, 30(4), 1993, pp. 317-320
Citations number
5
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
Journal title
ISSN journal
00261394
Volume
30
Issue
4
Year of publication
1993
Pages
317 - 320
Database
ISI
SICI code
0026-1394(1993)30:4<317:COAHCS>2.0.ZU;2-S
Abstract
We have measured the electrical and radiometric properties of a cryoge nic bolometer/amplifier which is intended for use as a transfer standa rd in an infrared spectral comparator facility under development at th e NIST. We have found its sensitivity to be approximately 20 pW, insta bility < 1 %, responsivity approximately 1,7 x 10(6) V/W, spatial nonu niformity < 1 % and nonlinearity < 1 % over five decades. These result s show that this device can be used for its intended purpose as an IR detector transfer standard.