RESPONSES OF LIF TLDS TO 10-40 KEV MONOENERGETIC PHOTONS FROM SYNCHROTRON-RADIATION

Citation
N. Nariyama et al., RESPONSES OF LIF TLDS TO 10-40 KEV MONOENERGETIC PHOTONS FROM SYNCHROTRON-RADIATION, Radiation protection dosimetry, 49(4), 1993, pp. 451-457
Citations number
NO
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
49
Issue
4
Year of publication
1993
Pages
451 - 457
Database
ISI
SICI code
0144-8420(1993)49:4<451:ROLTT1>2.0.ZU;2-F
Abstract
Glow curves of LiF TLDs (TLD-100) were measured for 10-40 keV monoener getic photons from synchrotron radiation, and the energy and dose resp onses were evaluated based on the glow curve data. The ratio of measur ed to calculated energy response was almost unity for the lower temper ature region of the glow curve including the main peak, and was about 2.0 for the higher one. Dose responses to 10 keV 30 keV X rays, and Co -60 gamma rays showed different supralinearity. The maximum Value of s upralinearity decreased with decreasing photon energy, that is, with i ncreasing the average linear energy transfer of incident photons.