IN-SITU THICKNESS CALIBRATION OF THIN TARGETS USED IN THE DETERMINATION OF STOPPING POWER

Citation
E. Unter et al., IN-SITU THICKNESS CALIBRATION OF THIN TARGETS USED IN THE DETERMINATION OF STOPPING POWER, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 83(4), 1993, pp. 577-580
Citations number
11
Categorie Soggetti
Physics, Nuclear","Nuclear Sciences & Tecnology","Instument & Instrumentation
ISSN journal
0168583X
Volume
83
Issue
4
Year of publication
1993
Pages
577 - 580
Database
ISI
SICI code
0168-583X(1993)83:4<577:ITCOTT>2.0.ZU;2-X
Abstract
We present a simple device, which allows us to determine the areal mas s density of thin targets in UHV with an accuracy better than 3%. It i s a modification of the principle used by Andersen et al. [Nucl. Instr . and Meth. 149 (1978) 137]. We compare the Rutherford backscattering rate from a reference target to that of the target under investigation . Both targets are alternately exposed to the beam. As a reference tar get we use a polished cone made of commercial aluminum alloy, which ro tates at constant speed upstream of the measured target. In this way, we can obtain absolute stopping cross sections of highly reactive targ ets, produced in-situ by evaporation.