Jm. Carreira et al., COMPUTERIZED AUTOMATIC SYSTEM TO OBTAIN INSPIRATION EXPIRATION CHEST ROENTGENOGRAMS WITHOUT PATIENT COLLABORATION/, Computer methods and programs in biomedicine, 52(3), 1997, pp. 185-193
We have developed a computerized automatic X-ray image acquisition sys
tem; It consists of conventional pressure and flow transducers connect
ed to a commercial data acquisition card installed in a PC/AT, the car
d output being connected through a relay to a conventional X-ray syste
m. Signal processing, that has been applied for automatic detection of
inspiration and expiration in respiratory signals, allows the X-ray e
quipment activation at a given moment of the respiratory cycle. Radiog
raphs are thus obtained either in inspiration or expiration at tidal v
olume on anaestesized experimental animals. The system has also been t
ested on non-collaborating patients. (C) 1997 Elsevier Science Ireland
.