Dj. Callans et al., EFFECT OF RATE AND COUPLING INTERVAL ON ENDOCARDIAL-R WAVE AMPLITUDE VARIABILITY IN PERMANENT VENTRICULAR SENSING LEAD SYSTEMS, Journal of the American College of Cardiology, 22(3), 1993, pp. 746-750
Objectives. We have observed sensing errors in third generation implan
table cardioverter-defibrillators that appear to be caused by variatio
n in the R wave amplitude during sinus rhythm, particularly after prem
ature beats. The purpose of this study was to quantify spontaneous R w
ave variability during sinus rhythm and to determine whether abrupt ch
anges in cycle length further augment R wave amplitude variability. Ba
ckground. Pacemaker sensing algorithms presume a relatively constant R
wave signal to establish a sensing threshold. The concept of a fixed
sensing threshold is not as applicable in third-generation cardioverte
r-defibrillators, which depend on automatic gain amplifiers to rapidly
detect ventricular fibrillation. These devices may be susceptible to
undersensing during sinus rhythm if significant variability in R wave
signal characteristics occurs. Methods. Twelve patients with combinati
on bradycardia pacing cardioverter-defibrillators were studied. The de
vice used (Cadence, Ventritex) allowed recording of real time, telemet
ered electrograms from the sensing lead system. Measurements were made
of the maximal range of the R wave amplitude during sinus rhythm and
in response to abrupt changes in heart rate produced by premature atri
al and ventricular stimuli. Results. The maximal range in R wave ampli
tude during sinus rhythm was 1.7 +/- 1.3 mV, or 23.7 +/- 19.2% of the
mean R wave amplitude. The R wave amplitude variability increased with
abrupt changes in cycle length, with a range of 2.8 +/- 1.5 mV, or 38
.8 +/- 18.3% of the mean R wave amplitude (p < 0.05 compared with sinu
s rhythm). In most patients, R wave amplitude and coupling interval de
monstrated an inverse proportional relation. Conclusions. There is sub
stantial variability in the R wave amplitude during sinus rhythm measu
red by permanent ventricular sensing lead systems, and this variabilit
y is further augmented by abrupt changes in cycle length. This phenome
non may explain the occurrence of undersensing of sinus rhythm in impl
antable cardioverter-defibrillators with automatic gain sense amplifie
rs.