P. Ostrega et al., SIMULTANEOUS DETERMINATION OF CD AND PB BY ATOMIC-ABSORPTION SPECTROMETRY WITH A TUNGSTEN-COIL ATOMIZER, Chemia Analityczna, 38(6), 1993, pp. 779-788
A simple, electrothermal atomizer with a the tungsten-coil was used fo
r simultaneous determination of Cd and Pb in natural water samples. Th
e construction of the low-voltage metal coil atomizer and the influenc
e of applied power, gas flow rate as well as other experimental condit
ions on the signal parameters were described in detail. The atomizatio
n behaviour of Cd and Pb in the presence of different matrices was als
o studied. Under optimal operating conditions the characteristic masse
s calculated from the peak-area (for 0.0044 A s), for Cd and Pb, were
0.8 and 1.8 pg respectively.