OPTIMIZATION OF FAST PROTEIN SEPARATIONS ON NONPOROUS SILICA-BASED STRONG ANION-EXCHANGERS

Citation
G. Jilge et al., OPTIMIZATION OF FAST PROTEIN SEPARATIONS ON NONPOROUS SILICA-BASED STRONG ANION-EXCHANGERS, Chromatographia, 37(11-12), 1993, pp. 603-607
Citations number
13
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00095893
Volume
37
Issue
11-12
Year of publication
1993
Pages
603 - 607
Database
ISI
SICI code
0009-5893(1993)37:11-12<603:OOFPSO>2.0.ZU;2-R
Abstract
The adsorbed coating technology using various vinylpyrrolidone-vinylim idazole copolymer compositions was carried out on 1.7 mum non-porous m onodisperse silica. It was shown that the retention properties and the loading capacity for bovine serum albumin (BSA) increases with the am ount of vinylimidazole in the copolymer composition. The retention beh avior of various proteins as a function of the salt composition in the eluent has been applied to find the optimal conditions for the synthe sis of the anion-exchange stationary phase. The suitability of these s upports for the fast separation of biological molecules is demonstrate d. The best resolution and the highest speed in protein analysis was o btained by using an anion exchanger of low capacity.